Trigger Coupling (Coupling) mode selection trigger signal to trigger circuit coupled in different ways, the purpose is to trigger a stable and reliable. Here are a few commonly used. AC-coupling is also called capacitive coupling. It only allows the exchange of components of the trigger signal trigger, trigger signal DC component has been cut off. DC component is usually not considered when the use of this coupling in order to form a stable trigger. However, if the trigger signal frequency is less than 10Hz, would cause trigger problems. DC-coupled (DC) does not cut off the DC component of the trigger signal. When the trigger signal or trigger signal the frequency of the lower duty cycle is large, the use of DC-coupled better. low-frequency suppression (LFR) is triggered when the trigger signal through the high-pass filter added to the trigger circuit to trigger low-frequency signal component is suppressed; high-frequency suppression (HFR) is triggered when the trigger signal through the low-pass filter added to the trigger circuit, trigger signal high-frequency components was inhibited. There are also synchronization for television repair television (TV) is triggered. Trigger coupling of these have their own scope of application is required to experience in use. 3. Trigger level (Level) and the trigger polarity (Slope) trigger level adjustment is also called the simultaneous regulation, which makes scanning and measured signal synchronization. Level adjustment knob adjustment trigger signal of the trigger level. Once the trigger signal exceeds a trigger set by knob usual, scan Jibei trigger. Clockwise rotation knob, trigger level rise; counterclockwise rotation knob, trigger level down. When the power level knob transferred to level the locked position, the trigger level automatically maintained within the range of the trigger signal, without level regulation can generate a stable trigger. When the signal waveform complex, a stable power level knob can not be triggered,With the release inhibition (Hold Off) waveforms of the release knob adjustment suppression time (scan pause time), can scan and waveform stability of synchronization. polarity switch is used to select the polarity of the trigger signal. Dial the "+" position when the direction of the signal increases, when the trigger signal is generated more than usual Trigger trigger. Dial in the "-" position when, in the direction of the signal to reduce, when the trigger signal is generated more than usual Trigger trigger. Trigger polarity and trigger level trigger signal jointly determined trigger point. 2.6 scan mode (SweepMode) scan automatic (Auto), normal (Norm) and single (Single) three kinds of scan mode. Auto: When no triggering signal input, or when the trigger signal frequencies below 50Hz, scan for the self-excited mode. Normal: When no trigger signal input, the scan at the preparatory level, there is no scan lines. After the arrival of trigger signals to trigger scan. Single: Single-button similar to the reset switch. Single-scan mode, press a single button to reset the circuit when the scan at this time ready (Ready) lights. Trigger signal generated from the arrival of a scan. After a single scan in preparation for lights eliminate. A single scan for the observation of non-periodic signal, or a single transient signals often need to take pictures of the waveform. outlined above describes the basic functions and operation of the oscilloscope. Oscilloscope there are other more complex functions, such as the delayed scanning, triggering delays, XY working methods and so on, is not presented here. Oscilloscope entry operation is easy, really skilled in the application will have to master. It is worth noting that, although the function of the oscilloscope more, but in many cases other instruments, meters and better. For example, in digital circuit experiment, a pulse width to determine whether the narrow single pulse occurs, with the logic of simple multi-pen; measurement of single pulse width, used logic analyzer better.1. The acquisition of baseline: When the operator instructions in the use of non-use of the oscilloscope, the first to get one of the most detailed level of the baseline, and then the probe can be used for other measurements, the specific methods are as follows: (1) preset panel of switches, knobs. set moderate intensity, focus and support focus set is moderate, the vertical input coupling set "AC,,, the vertical voltage range selection set" 5mv/div ", select the vertical methods of work home" CHl ", fine-tune the calibration of the vertical position of the sensitivity set" CAL ", the vertical channel sync source selection set the middle position, the vertical position of the middle set, A and B together with pre-scan time-factor in the" 0.5ms/div ", A scan time fine-tuning the calibration set position" CAL'', horizontal displacement of set intermediate position, scanning work home "A", triggering synchronous way home "AUTO", the slope of switch set "+"